Instrument Features:
Powerful Functions:
Rich Electrical Interface:
Easy to Interact
High precision/high resolution
●Adopt borosilicate hard glass burette with low expansion coefficient and low temperature volume error to achieve high-precision infusion of 1/20000 burette volume, and the infusion accuracy is as high as ±10 μL (10 mL), which is better than the relevant JJG 814 and ISO 8655-2 standards. Require
●Based on a highly integrated ceramic-based measurement and control circuit, digital photometric, temperature, and potential electrodes are designed to realize in-situ photothermal signal measurement
●Using high-frequency sampling data filtering algorithm, double-precision floating-point operation mode, and high-speed operation processor to ensure ultra-high effective data bits and accurate and reliable results
●Adopt advanced neural network, PID control and other algorithms to precisely control the end point infusion process and obtain the most accurate end point value
Audit Trail
●It can meet the relevant requirements of GXP, FDA 21 CFR Part 11 and other laws and regulations without a computer, and realize the whole process of audit trail
●With password aging management function, the operator’s identity can be confirmed through password login or biometric identification. Operators with different identities have corresponding levels of operation authority. Actions and comments for each step on the titrator are automatically stored. It can be displayed and queried in stages in the work log
●Each measurement data will be permanently saved, all data-related information, such as: operator, operation time, experimental process, equipment used, etc., can be traced back, and the modification of data is also recorded.
Accuracy |
±0.1% |
Precision |
≤0.1% |
Burette installation method |
Tool-free replacement |
Burette resolution |
1/20000 |
Burette accuracy |
±10μL(10 mL) |
Burette addition speed |
1~99 mL /min |
Measuring range |
±2400 mV / ±20Ph |
Resolution |
0.01 mV / 0.001 pH |
Indication error |
±0.03 %FS / 0.005 pH |
Indication repeatability |
≤0.25 % / 0.002 pH |
Input current |
≤1×10-12 A |
Input impedance |
≥3×1012 Ω |
Temperature and humidity measurement range |
0~125℃,10~85%RH |
Temperature and humidity resolution |
0.1℃,1% RH |
Temperature and humidity measurement error |
±0.3℃,±5% RH |
Titration mode |
Dynamic titration, equivalent titration, endpoint titration, manual titration |
Measurement method |
pH, Potential, Ion Concentration, Temperature |