Core Description
The instrument has standard four phase voltage and three-phase current output (six phase voltage and six phase current output). It can not only test various traditional relays and protection devices, but also test various modern microcomputer protection, especially for transformer differential power protection and standby automatic switching device. The test is more convenient and perfect.
3*20A |
|||
Single phase current output (effective value) |
0 — 20A / phase, |
accuracy |
0.2% ±5mA |
Three phase parallel output (effective value) |
0 — 60A / three-phase in-phase parallel output |
||
Allowable working value of phase current for a long time (effective value) |
10A |
||
Maximum output power of each phase |
200va |
||
Maximum output power of three-phase parallel current |
600VA |
||
Maximum allowable working time of three parallel current output |
30s |
||
Frequency range |
0 — 1000Hz |
accuracy |
0.01Hz |
Harmonic frequency |
2-20 times |
||
Phase |
0 — 360 ° |
accuracy |
0.1 ° |
3*30A |
|||
Single phase current output (effective value) |
0 — 30A / phase, |
accuracy |
0.2% ±5mA |
Three phase parallel output (effective value) |
0 — 90a / three-phase in-phase parallel output |
||
Allowable working value of phase current for a long time (effective value) |
10A |
||
Maximum output power of each phase |
300VA |
||
Maximum output power of three-phase parallel current |
800VA |
||
Maximum allowable working time of three parallel current output |
30s |
||
Frequency range |
0 — 1000Hz |
accuracy |
0.01Hz |
Harmonic frequency |
2-20 times |
||
Phase |
0 — 360 ° |
accuracy |
0.1 ° |
3*30A |
|||
Single phase current output (effective value) |
0 — 40A / phase |
accuracy |
0.2% ±5mA |
Three phase parallel output (effective value) |
0 — 120a / three-phase in-phase parallel output |
||
Allowable working value of phase current for a long time (effective value) |
10A |
||
Maximum output power of each phase |
420va |
||
Maximum output power of three-phase parallel current |
1000VA |
||
Maximum allowable working time of three parallel current output |
10s |
||
Frequency range |
0 — 1000Hz |
accuracy |
0.01Hz |
Harmonic frequency |
2-20 times |
||
Phase |
0 — 360 ° |
accuracy |
0.1 ° |
6*20A |
|||
Single phase current output (effective value) |
0 — 20A / phase |
accuracy |
0.2% ±5mA |
Three phase parallel output (effective value) |
0 — 120a / six same phase parallel output |
||
Allowable working value of phase current for a long time (effective value) |
10A |
||
Maximum output power of each phase |
200va |
||
Maximum output power of three-phase parallel current |
800VA |
||
Maximum allowable working time of three parallel current output |
30s |
||
Frequency range |
0 — 1000Hz |
accuracy |
0.01Hz |
Harmonic frequency |
2-20 times |
||
Phase |
0 — 360 ° |
accuracy |
0.1 ° |
6*30A |
|||
Single phase current output (effective value) |
0 — 30A / phase |
accuracy |
0.2% ±5mA |
Three phase parallel output (effective value) |
0 — 180A / six same phase parallel output |
||
Allowable working value of phase current for a long time (effective value) |
10A |
||
Maximum output power of each phase |
300VA |
||
Maximum output power of three-phase parallel current |
1000VA |
||
Maximum allowable working time of three parallel current output |
30s |
||
Frequency range |
0 — 1000Hz |
accuracy |
0.01Hz |
Harmonic frequency |
2-20 times |
||
Phase |
0 — 360 ° |
accuracy |
0.1 ° |
DC current source
DC current output 0 – ± 10A / phase, accuracy |
0.2% ±5mA |
AC Voltage Source
Single phase voltage output |
(effective value) 0 — 125V / phase |
accuracy |
0.2% ±5mV |
Line voltage output (effective value) |
0 — 250V |
||
Phase voltage / line voltage output power |
75va / 100VA |
||
Frequency range |
0 — 1000Hz |
accuracy |
0.001Hz |
Harmonic frequency |
2-20 times |
||
Phase |
0 — 360 ° |
accuracy |
0.1 ° |
DC voltage source
Single phase voltage output amplitude |
0 — ± 150V |
accuracy |
0.2% ±5mV |
Line voltage output amplitude |
0 — ± 300V |
||
Phase voltage / line voltage output power |
90va / 180va |
Switching value terminal
Switching value input terminal |
8 pairs |
Empty contact |
1 — 20mA, 24V, internal active output of the device |
Potential reversal |
passive contact:low resistance short circuit signal |
Active contact |
0-250V DC |
Switching value output terminal |
4 pairs, empty contact, breaking capacity:110V / 2a, 220V / 1A |
Other
Time range |
1ms — 9999s, measurement accuracy 1ms |
Unit volume and weight |
410 x 190 x 420mm3, about 18kg |
Power Supply |
AC220V±10%,50Hz,10A |